Video S1 Cadaver A, right lower lobe. Air leakage from the staple line using GST®60D was observed at an airway pressure of 30 cmH2O, whereas no air leakage was observed at the same pressure when using GST®60D with SLR.
Video S2 Cadaver A, right middle lobe. Air leakage from the staple line using GST®60D with SLR was observed at an airway pressure of 45 cmH2O.
Video S3 Cadaver B, right lower lobe. Air leakage was observed from a portion of the lung parenchyma unrelated to the staple line at an airway pressure of 30 cmH2O.
Video S4 Cadaver B, right middle lobe. Air leakage was first observed at an airway pressure of 20 cmH2O from a portion of the lung parenchyma unrelated to the staple line.
Video S5 Cadaver C, left lower lobe. Air leakage from the staple line using GST®60D was observed at an airway pressure of 25 cmH2O, whereas no air leakage was observed at the same pressure when using GST®60D with SLR.
Video S6 Cadaver C, left lower lobe. An additional resection, including the two prior staple lines, was performed using GST®60D with SLR.