The motion of intra-thoracic tumors and radiosensitive normal organs is an important consideration when planning curative radiotherapy. An individualized assessment of such motion is preferable over use of standard 'planning margins', which may be based upon tumor location or patient characteristics ( 1-3). This is primarily because many tumors move less than 1 cm in any direction ( 3), thereby requiring smaller treatment fields. In addition, lung tumors may exhibit complex motions patterns, including hysteresis and be influenced by cardiac motion ( 4). Furthermore, patients with locally advanced lung cancer may exhibit motion in metastatic mediastinal nodes that exceeds that of the primary tumor or move maximally in a different phase ( 2, 5). Identifying such variations in tumor motion can minimize the risk of compromising the effectiveness of radiotherapy delivery. Based on these studies, the European Organization for Research and Treatment of Cancer guidelines recommend use of respiration-correlateDCT or 4-Dimensional CT (4DCT) scans for planning ( 6).
As 4DCT allows for the motion trajectory of a tumor to be imaged, it reduces the risk of introducing systematic errors into radiotherapy planning ( 7). Individualized 4DCT planning is particularly important for planning high-precision SABR in early-stage NSCLC ( 8). The widespread implementation of SABR in the Netherlands was accompanied by quality assurance programs for 4DCT imaging ( 9), and attention to such detail may have contributed to the remarkable survival improvements observed in the entire Netherlands population for Stage I lung cancer patients aged 75 years and older, who were diagnosed between 2001 and 2009 ( 10). The two-year survival in elderly Dutch patients undergoing radiotherapy improved from 35.8% to 52.5%, and median survival increased from 16.4 to 24.4 months.
Although multi-slice 4DCT scanners have been commercially available since 2003, this facility is not universally available and clinicians have had to explore alternative approaches. In this issue, Shen et al. describe an approach utilizing breath-hold CT scans at the end of coached 'regular and quiet respiration' ( 11). Although this approach captures more motion than a single conventional CT acquired at a random phase of respiration, some key drawbacks must be kept in mind. Firstly, patients who present for SABR are frequently unfit for surgery due to compromised pulmonary function ( 12). The ability of all such patients to perform a breath-hold procedure reproducibly is questionable. Furthermore, previous work has shown that audio-coaching can both increase the amplitude of tumor motion, as well as displace tumors ( 13). Therefore, the same audio-coaching must be performed during the delivery of SABR, where this may lead to longer treatment times. An alternative low-technology approach for acquiring 4D information is slow CT scanning, which uses a slow gantry rotation to acquire images of tumor position taken
at different points in the respiratory cycle ( 14). Dutch SABR
recommendations suggest that if only 3DCT scans are available,
the internal target volume for planning should be based either
on multiple slow CT-scans covering the full tumour trajectory,
or adding an additional 3-5 mm margin in all directions to the
CTV determined on a single slow CT-scan ( 15). An alternative
recommendation was to acquire a minimum of 3 conventional
rapid planning scans over the entire tumor trajectory. FDGPET
scans have also been proposed for this purpose ( 16),
although it is unclear if clinical outcomes using FDG-PET scans
alone (without 4DCT) will produce acceptable local control
rates. Additionally, the loss of contrast resolution has lead to
uncertainties regarding the optimal use of FDG-PET for this
purpose ( 17). Fluoroscopy in addition to a conventional CT
scan has also been shown to be suboptimal for SABR ( 18).
Although some studies suggest that a single 4DCT scan
acquired during quiet, uncoached respiration generates
reproducible internal target volumes ( 18-20), artifacts on 4DCT
have been reported in a significant proportion of such imaging
studies ( 21). The latter highlights the need for careful review of
images at the time of acquisition to ensure that the tumor region
is free of artifacts.
After tumor motion has been defined, the planning target
volume derived may be based on a motion-encompassing
internal target volume that can be derived from the union of
separate volumes in each phase of a 4DCT, or alternatively,
from contouring on a maximum intensity projection of the
4DCT dataset ( 15, 22). Dutch guidelines also accept use of
an alternative to the internal target volumes concept, which is
based on time-averaged mean positions of the tumor. Finally,
changes in anatomy and breathing pattern between 4DCT and
treatment, as well as between SABR fractions, can result in the
internal target volume used for planning not being appropriate
during treatment delivery ( 23). Therefore, even the use 4DCT to
derive personalized margins to account for tumor motion alone
may not be sufficient to ensure optimal cure rates. Daily image
guidance strategies to visualize treatment targets are important
during SABR to ensure that these individualized, reduced
margins provide adequate coverage ( 24).
|
References
- van Sörnsen de Koste JR, Lagerwaard FJ, Nijssen-Visser MR, Graveland
WJ, Senan S. Tumor location cannot predict the mobility of lung tumors:
a 3D analysis of data generated from multiple CT scans. Int J Radiat Oncol
Biol Phys 2003;56:348-54.[LinkOut]
- Pantarotto JR, Piet AH, Vincent A, van Sörnsen de Koste JR, Senan S.
Motion analysis of 100 mediastinal lymph nodes: potential pitfalls in
treatment planning and adaptive strategies. Int J Radiat Oncol Biol Phys
2009;74:1092-9.[LinkOut]
- Liu HH, Balter P, Tutt T, Choi B, Zhang J, Wang C, et al. Assessing
respiration-induced tumor motion and internal target volume using fourdimensional
computed tomography for radiotherapy of lung cancer. Int J
Radiat Oncol Biol Phys 2007;68:531-40.[LinkOut]
- Seppenwoolde Y, Shirato H, Kitamura K, Shimizu S, van Herk M, Lebesque
JV, et al. Precise and real-time measurement of 3D tumor motion in lung
due to breathing and heartbeat, measured during radiotherapy. Int J Radiat
Oncol Biol Phys 2002;53:822-34.[LinkOut]
- Donnelly ED, Parikh PJ, Lu W, Zhao T, Lechleiter K, Nystrom M, et al.
Assessment of intrafraction mediastinal and hilar lymph node movement
and comparison to lung tumor motion using four-dimensional CT. Int J
Radiat Oncol Biol Phys 2007;69:580-8.[LinkOut]
- De Ruysscher D, Faivre-Finn C, Nestle U, Hurkmans CW, Le Péchoux
C, Price A, et al. European Organisation for Research and Treatment of
Cancer recommendations for planning and delivery of high-dose, highprecision
radiotherapy for lung cancer. J Clin Oncol 2010;28:5301-10.[LinkOut]
- Slotman BJ, Lagerwaard FJ, Senan S. 4D imaging for target definition in
stereotactic radiotherapy for lung cancer. Acta Oncol 2006;45:966-72.[LinkOut]
- Underberg RW, Lagerwaard FJ, Cuijpers JP, Slotman BJ, van Sörnsen de
Koste JR, Senan S. Four-dimensional CT scans for treatment planning in
stereotactic radiotherapy for stage I lung cancer. Int J Radiat Oncol Biol
Phys 2004;60:1283-90.[LinkOut]
- Hurkmans CW, van Lieshout M, Schuring D, van Heumen MJ, Cuijpers
JP, Lagerwaard FJ, et al. Quality assurance of 4D-CT scan techniques
in multicenter phase III trial of surgery versus stereotactic radiotherapy
(radiosurgery or surgery for operable early stage (stage 1A) nonsmall-
cell lung cancer [ROSEL] study). Int J Radiat Oncol Biol Phys
2011;80:918-27.[LinkOut]
- Haasbeek CJ, Palma D, Visser O, Lagerwaard FJ, Slotman BJ, Senan S.
Survival Improvements in Elderly Patients presenting with Early Stage
Lung Cancer in the Netherlands between 2001 and 2009. Ann Oncol.
2012. [Epub ahead of print].
- Shen G, Wang YJ, Sheng HG, Duan XP, Wang JL, Zhang WJ, et al. Double CT imaging can measure the respiratory movement of small pulmonary tumors during stereotactic ablative radiotherapy. J Thorac Dis 2012;4:131-40.
- Haasbeek CJ, Lagerwaard FJ, Antonisse ME, Slotman BJ, Senan S. Stage I
nonsmall cell lung cancer in patients aged > or =75 years: outcomes after
stereotactic radiotherapy. Cancer 2010;116:406-14.[LinkOut]
- Haasbeek CJ, Spoelstra FO, Lagerwaard FJ, van Sörnsen de Koste JR,
Cuijpers JP, Slotman BJ, et al. Impact of audio-coaching on the position of
lung tumors. Int J Radiat Oncol Biol Phys 2008;71:1118-23.[LinkOut]
- Lagerwaard FJ, Van Sornsen de Koste JR, Nijssen-Visser MR, Schuchhard-
Schipper RH, Oei SS, Munne A, et al. Multiple "slow" CT scans for
incorporating lung tumor mobility in radiotherapy planning. Int J Radiat
Oncol Biol Phys 2001;51:932-7.[LinkOut]
- Hurkmans CW, Cuijpers JP, Lagerwaard FJ, Widder J, van der Heide
UA, Schuring D, et al. Recommendations for implementing stereotactic
radiotherapy in peripheral stage IA non-small cell lung cancer: report from
the Quality Assurance Working Party of the randomised phase III ROSEL
study. Radiat Oncol 2009;4:1.[LinkOut]
- Caldwell CB, Mah K, Skinner M, Danjoux CE. Can PET provide the
3D extent of tumor motion for individualized internal target volumes?
A phantom study of the limitations of CT and the promise of PET. Int J
Radiat Oncol Biol Phys 2003;55:1381-93.[LinkOut]
- Werner-Wasik M, Nelson AD, Choi W, Arai Y, Faulhaber PF, Kang P, et al.
What is the best way to contour lung tumors on PET scans? Multiobserver
validation of a gradient-based method using a NSCLC digital PET
phantom. Int J Radiat Oncol Biol Phys 2012;82:1164-71.[LinkOut]
- van der Geld YG, Senan S, van Sörnsen de Koste JR, van Tinteren H,
Slotman BJ, Underberg RW, et al. Evaluating mobility for radiotherapy
planning of lung tumors: a comparison of virtual fluoroscopy and 4DCT.
Lung Cancer 2006;53:31-7.[LinkOut]
- Haasbeek CJ, Lagerwaard FJ, Cuijpers JP, Slotman BJ, Senan S. Is adaptive
treatment planning required for stereotactic radiotherapy of stage I nonsmall-
cell lung cancer? Int J Radiat Oncol Biol Phys 2007;67:1370-4.[LinkOut]
- Guckenberger M, Wilbert J, Meyer J, Baier K, Richter A, Flentje M. Is a single respiratory correlated 4D-CT study sufficient for evaluation of breathing motion? Int J Radiat Oncol Biol Phys 2007;67:1352-9.[LinkOut]
- Yamamoto T, Langner U, Loo BW Jr, Shen J, Keall PJ. Retrospective
analysis of artifacts in four-dimensional CT images of 50 abdominal
and thoracic radiotherapy patients. Int J Radiat Oncol Biol Phys
2008;72:1250-8.[LinkOut]
- Underberg RW, Lagerwaard FJ, Slotman BJ, Cuijpers JP, Senan S. Use of
maximum intensity projections (MIP) for target volume generation in
4DCT scans for lung cancer. Int J Radiat Oncol Biol Phys 2005;63:253-60.[LinkOut]
- Purdie TG, Bissonnette JP, Franks K, Bezjak A, Payne D, Sie F, et al.
Cone-beam computed tomography for on-line image guidance of lung
stereotactic radiotherapy: localization, verification, and intrafraction tumor
position. Int J Radiat Oncol Biol Phys 2007;68:243-52.[LinkOut]
- Bissonnette JP, Purdie TG, Higgins JA, Li W, Bezjak A. Cone-beam
computed tomographic image guidance for lung cancer radiation therapy.
Int J Radiat Oncol Biol Phys 2009;73:927-34.[LinkOut]
Cite this article as: Senthi S, Senan S. Imaging for high-precision
thoracic radiotherapy. J Thorac Dis 2012;4(2):106-108. doi: 10.3978/
j.issn.2072-1439.2012.03.07
|