Fatal bleeding due to sternal steel wire fracture following open-heart surgery
Case Report

Fatal bleeding due to sternal steel wire fracture following open-heart surgery

Wenhui Gong, Xiaofeng Ye, Zhe Wang, Sen Li, Qiang Zhao

Department of Cardiovascular Surgery, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai 200025, China

Correspondence to: Qiang Zhao, MD. Department of Cardiovascular Surgery, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, 197 Ruijin Er Road, Shanghai 200025, China. Email: zq11607@rjh.com.cn.

Abstract: Nowadays, stainless steel wires suturing is the most common and safety way for the median sternotomy closure. And it is also accepted by our center. Albeit, in July 2014, a patient who underwent left ventricular outflow tract plus mitral valvuloplasty in extracorporeal circulation suffered a fatal bleeding due to fractured steel wire. The fractured steel wire punctured the free wall of right ventricle and left a 3 mm diameter wound, thus caused the 1,000 mL drainage within half an hour. In our opinion, it was the instant increase in yield strength after strong cough that makes the steel wire fractured.

Keywords: Heart surgery; hemorrhage; complication


Submitted Apr 25, 2015. Accepted for publication Sep 17, 2015.

doi: 10.3978/j.issn.2072-1439.2015.10.52


Introduction

Postoperative hemorrhage is a common complication following open heart surgery. It has been reported that the incidence of chest re-exploration ranges from 2% to 6.2%. However, massive bleeding in a short period after surgery is not frequent indeed. Here is the case of serious bleeding due to cardiac puncture caused by the fractured ends of the steel wire.


Case report

A 60-year-old man with hypertrophic obstructive cardiomyopathy and severe mitral regurgitation underwent ventricular septal myectomy and mitral valve repair in July 2014. And we closed the sternum by using No. 6 steel wires in a figure-of-eight suturing manner. On the first day after operation, the patient status was stable and the drainage volume in the first 24 hours was 260 mL. Next day, the patient was transferred back from intensive care unit (ICU) to telemetry ward. In the morning of third postoperative day, the patient woke up with serious cough. Suddenly, the drainage volume increase to about 1,000 mL within half an hour. Due to massive bleeding from chest, the patient developed hypovolemic shock and being rushed right to the operation room for re-exploration. Exploring the chest, we found that the upper two steel wires were intact, but the lowest wire was broken into two segments (Figure 1). It was the fractured steel wire that lacerate the free wall of the right ventricle and resulted in about 3 mm damage. We stopped the bleeding successfully by using interrupted 4-0 prolene pledgeted horizontal mattress stitches. After 2 days later, the patient was back to ward and discharged home on the 8th postoperative day with good condition. We carefully examined the end of fractured steel wire and found its cross-section was neat and smooth, which was different from the man-made cut (Figure 2).

Figure 1 Schematic of wire fracture site. The lower left black arrow indicates the site of the wire fractured, and the upper right grey arrow indicates the site where the wire was cut.
Figure 2 The cross-section of the fractured and the man-made cut steel wires. The left wires show the sharp man-made cut section; the right fractured wires show the neat and smooth cross-section due to sudden stress increases.

Discussion

Actually, there are many reasons for excessive bleeding after cardiac surgery, such as blood coagulation disorders after extracorporeal circulation, preoperative multiple organ dysfunctions, intraoperative hemostasis, and protamine supplement deficiency (1). However, we have never witnesses re-exploration caused by fatal bleeding due to fractured steel wire.

Mainly, two reasons respond for the sternal wire fracture: the overloading due to the patient’s activity and the wire mechanical fatigue, such as growth of sternum, activities, and the sharp increase of wires’ yield strength (2). Usually, wire fracture occurs a week or more after surgery. However, in this case, the wire fractured on the second postoperative day. What’s more, the fractured point was under the sternum rather than the usual screwed location (Figure 1). Stable sternal fixation is the sine qua non for healing of median sternotomy. So new techniques and materials for median sternotomy closure are needed which are less prone to cause mechanical fatigue.

A large cohort of consecutive cardiac operations found that sternal closure with figure-of-eight single wires was better than simple interrupted wires (3). And also in our center, usually we close sternum with steel wires in the shape of figure-of-eight. Nevertheless, the optimal technique for sterna closure remains controversial. It has been reported that as for sternal fixation, double wire loops is better than single wire (4). Because of the better biomechanical stability, it has been accepted by many centers. And also the complication presented may not happen if double wire closure was used. There are many other median sternotomy closure techniques too. A retrospective study indicates that the sternal plating is an effective option to treatment of sternal wound dehiscence associated with sternal instability (5).


Conclusions

The wire rupture in present case was the result of not being able to afford the instant yield strength after severe coughing. New techniques and materials for median sternotomy closure, external chest strap, combined with appropriate postoperative respiratory therapy can minimize the occurrence of such complications.


Acknowledgements

None.


Footnote

Conflicts of Interest: The authors have no conflicts of interest to declare.


References

  1. Charalambous CP, Zipitis CS, Keenan DJ. Chest reexploration in the intensive care unit after cardiac surgery: a safe alternative to returning to the operating theater. Ann Thorac Surg 2006;81:191-4. [PubMed]
  2. Chao J, Voces R, Peña C. Failure analysis of the fractured wires in sternal perichronal loops. J Mech Behav Biomed Mater 2011;4:1004-10. [PubMed]
  3. Almdahl SM, Halvorsen P, Veel T, et al. Avoidance of noninfectious sternal dehiscence: figure-of-8 wiring is superior to straight wire closure. Scand Cardiovasc J 2013;47:247-50. [PubMed]
  4. Losanoff JE, Basson MD, Gruber SA, et al. Single wire versus double wire loops for median sternotomy closure: experimental biomechanical study using a human cadaveric model. Ann Thorac Surg 2007;84:1288-93. [PubMed]
  5. Fawzy H, Osei-Tutu K, Errett L, et al. Sternal plate fixation for sternal wound reconstruction: initial experience (retrospective study). J Cardiothorac Surg 2011;6:63. [PubMed]
Cite this article as: Gong W, Ye X, Wang Z, Li S, Zhao Q. Fatal bleeding due to sternal steel wire fracture following open-heart surgery. J Thorac Dis 2015;7(10):E517-E519. doi: 10.3978/j.issn.2072-1439.2015.10.52

Download Citation